Chlorine Electrochemical Sensor EC-4CL2-50
EC-4CL2-50 is CL2 sensor which detects chlorine in level from 0~50ppm.Since the output current is fairly large, it can detect the low concentration of Cl2 with great accuracy.
Electrochemical Chlorine Sensor EC-4Cl2-50 has very excellent repeatability, long term stability, selectivity and resistivity, it is useful for the concentration control of Cl2 at water sterilization.
The Chlorine(Cl2) Electrochemical Sensor EC-4Cl2-50 can be retrofitted easily to existing product designs.
Dimensions in millimeters(±0.1mm)
Target Gas and Concentration
Gas | P/N | Operating principle | Detection Range | Maximum Overload |
Chlorine (Cl2) | SEC-4Cl2-M | 3-electrode electrochemical | 0~50ppm | 100ppm |
Gas Sensitivity Characteristics
Gas | Concentration & Output Current |
Chlorine(Cl2) | 0.40 ± 0.2μA /ppm |
Ratings & environment
Response Time (T90) | ≤30S (Typical 19S) |
Repeatability | <±2% signal |
Linearity | Linear |
Long term output Drift | <2% signal/month |
Resolution | 0.1 ppm |
Recommended Load | 5~30 Ω |
Bias Voltage | 0 mV |
Operating Temp. Range | -20℃~ 50℃ |
Operating Humidity | 15% RH ~ 90% RH(non-condensing) |
Operating Pressure Range | 800 ~ 1200 mbar |
Storage & Shelf Life
Recommended Storage Temperature | 0℃~ +20℃ in sealed package |
Expected Operating Life | 24 months in air |
Storage Life | 6 months in original package |
Standard Warranty | 18 months from the date of delivery |
Cross Sensitivity Data
Gas | Concentration (ppm) | Output signal (ppm Cl2 equivalent) |
Hydrogen | 4000 | 0 |
Hydrogen Sulfide | 15 | -3 |
Sulfur Dioxide | 10 | 0 |
Nitric oxide | 50 | 0 |
Nitriogen Dioxide | 5 | 6 |
Ammonia | 100 | 0 |
Carbon Dioxide | 10000 | 0 |
Carbon Monoxide | 200 | 0 |
Temperature Data
Sensitivity Temperature Dependence